RF Measurement — Broadband power meter
PM-1
RF power, read in dBm or mW.
The PM-1 is a 10 – 1000 MHz broadband RF power meter, from 0.1 mW to 2 W (−10 to +33 dBm), accurate to ±0.2 dB. Relative mode to measure a cable loss or an amplifier gain, ±20 dB compensation, threshold alarms. 9 V battery, aluminium housing, 200 g.
Bandwidth
10 – 1000 MHz
Range
−10 to +33 dBm · 0.1 mW – 2 W
Accuracy
±0.2 dB typ.
Relative mode
Loss and gain
Power
9 V battery · 8 h
Highlights
Power in dBm or mW
CW power measurement from 0.1 mW to 2 W on a 4-digit 7-segment display. ±0.2 dB typical accuracy from 0 to 33 dBm, ±0.4 dB below. CW, QPSK (digital mics) and pulsed FM modes.
−10 to +33 dBm · 10 – 1000 MHz
Relative mode
A 1 s press stores the reference; the PM-1 then shows the difference in dB: insertion loss of a cable, a filter, a splitter, or the gain of an amplifier.
Reference · reading in dB
±20 dB compensation
Enter the value of an external attenuator to measure beyond 2 W, of a directional coupler to monitor forward or reflected power, or of an amplifier to go below 0.1 mW.
Attenuator · coupler · amplifier
Alarms
Adjustable high and low thresholds: buzzer and open-drain output (100 mA) to monitor a transmitter continuously. Overload alarm above 2 W.
High / low thresholds · buzzer · output
9 V battery, aluminium housing
Up to 8 h continuous on a 9 V alkaline battery. Supplied with a reference BNC cord, a BNC-SMA adapter and battery. (*no battery included for shipping outside europe)
84 × 59 × 31 mm · 200 g · 8 h
In detail
The PM-1 is a broadband RF power meter. It measures CW power from 10 to 1000 MHz, from 0.1 mW to 2 W, and displays it in mW or dBm. CW, QPSK (digital mics) and pulsed FM modes.
The relative mode stores a reference and shows the difference in dB: insertion loss of a coax cable, a filter or a splitter, gain of an amplifier. With a pocket transmitter or an IEM as the source, a cable is tested in seconds.
A −20 to +20 dB compensation accounts for an external attenuator, to measure beyond 2 W, a directional coupler, to monitor the forward or reflected power of a transmitter in service, or an amplifier, to go below 0.1 mW.
High and low alarm thresholds trigger the buzzer and an open-drain output (100 mA), for a warning or to switch an amplifier. An alarm flags an input above 2 W.
Aluminium housing, 50 Ω BNC, 9 V alkaline battery for up to 8 h of measurement. Supplied with a reference BNC cord, a BNC-SMA adapter and the battery. Accessories: adapters, coupler, 6 dB / 10 W attenuator, rear panel with external power input, 1U panel for four PM-1.
Specifications
The numbers that matter.
Condensed excerpt. See the datasheet for detailed conditions.
Measurement
Functions
Power
Mechanical
Use cases
Transmitters
Check a transmitter's power
Pocket transmitter, rack IEM or walkie: direct reading in dBm or mW, up to 2 W. Beyond that, an attenuator and the compensation.
Cables
Test a coax cable
Reference on the source, then insert the cable under test: the loss reads in dB. Compared with the cable's theoretical loss, it reveals a damaged connector or cable.
Monitoring
Monitor a transmitter in service
With a directional coupler, the PM-1 reads forward or reflected power while transmitting; the alarm thresholds warn of a drift.